Deskripsi Alat |
FACSLyric 3L10C Instrument IVD
Merek: BD Biosciences
Optics
Available system configurations
4-color: 2-laser (blue, red) (3-1) 6-color: 2-laser (blue, red) (4-2)
8-color: 3-laser (blue, red, violet) (4-2-2) 10-color: 3-laser (blue, red, violet) (4-3-3) 12-color: 3-laser (blue, red, violet) (4-3-5)
Solid-state laser specifications
Blue laser: 488 nm, 20 mw Red laser: 640 nm, 40 mw Violet laser: 405 nm, 40 mw
Beam spot size (all lasers)
9 m x 63 m
Optical alignment
Auto alignment on demand
Flow-cell lens 1.2 NA
FSC detector
Photodiode
SSC and FL detectors
PMT
See filter guide for optical configurations.
Fluidics
Flow cell
Stainless steel with low coefficient of thermal expansion for predictable, stable performance Cuvette internal cross-section
430 m x 180 m
Sample flow rates
Low: 12 L/min
Medium: 60 L/min
High: 120 L/min
High sensitivity: 50 L/min
Fluid capacity Standard 5-L tanks Optional 10-L tanks
Adapter available for 20-L BD FACSFlowTM cubitainer
Sheath core stream fluid velocity Normal: 5.4 m/s
High sensitivity: 2.7 m/s
Sheath fluid consumption
Normal: 13.6 mL/min
High sensitivity: 6.6 mL/min
Supported tubes, plates and tube racks
• With BD FACS Universal Loader Tubes 30-tube rack (12 x 75-mm tubes) 40-tube rack (12 x 75-mm tubes)
Plates
96 Falcon® standard height, round, polystyrene 96 Falcon® standard height, flat, polystyrene 96 Falcon® standard height, round, polypropylene
96 Falcon® standard height, conical, polypropylene
384 Greiner standard height, flat, polystyrene
96 Falcon®, half deep, conical, polypropylene
96 Falcon®, deep, conical, polypropylene 96 Milipore, filter bottom, polypropylene
• With manual tube port
Falcon® 5 mL (12 x 75-mm) polystyrene and polypropylene
BD TrucountTM 5 mL (12 x 75 mm) Falcon 15 mL
Falcon 50 mL
Microcentrifuge 2 mL Sample dead volume
30 μL (12 x 75-mm tubes)
Cytometer schedule settings
Pre-programmed startup and idle shutdown
Software
• Integrated bi-directional LIS interface using BD FACSLinkTM software
• Support for 21 CFR Part 11 workflow with audit trail and e-signature
• Universal setup for fast and convenient instrument setup and standardization
• Single-tube QC with BDTM CS&T beads • QC module with Levey-Jennings plots
• Two applications
BD FACSuite Application
• User-defined assays
• User-defined plots
• User-defined worksheets and reports • User-defined tube/reference settings • Expression editing
BD FACSuite Clinical Application
Pre-configured workflow and pre-set templates for the following BD IVD assays:
• BD TritestTM
• BD MultitestTM 4-Color BD Multitest 6-Color TBNK • BD TrucountTM control
• Report in 26 languages
QC
Automated single-tube QC with BDTM CS&T beads
Performance
Acquisition rate
Up to 35,000 events per second. No limit on number of events acquired in a single FCS file
Carryover
<0.10% with default SIT flush <0.05% with 3 or more SIT flushes Sensitivity
FITC: <85 MESF
PE: <20 MESF
Channel Qr* (x1,000)
FITC 20
PE 133
PerCP-CyTM5.5 13 PE-CyTM7 17
APC 10
BD HorizonTM APC-R700 8 APC-CyTM7 7
BD Horizon V450 47
BD Horizon V500 17 BD Horizon BV605 133
BD Horizon BV711 43
BD Horizon BV786 16
Fluorescence precision
<3% CV for chicken erythrocyte nuclei (CEN)
Fluorescence linearity
2 ±0.05% for CEN
Data resolution
Uncompensated data has a range of 0–262,143
SSC and FSC resolution
Enables separation of 0.2-μm beads from noise
System throughput
50 minutes for a 40-tube rack with a standard BD Tritest assay stopping rule on samples with normal CD4 counts (approximately 1190 cells/μl).
40 minutes for a 96-well plate, using default mix settings, a two-second asquisition, and a SIT flush in between each well and no preview before acquring or report review delay.
Parameters
Area (A), Width (W), Height (H) for all channels and Time (T). Total of 43 parameters available.
3 Scales:
Linear (A, W, H), Logarithmic (A, H) Biexponential (A, W, H)
Compensation
Full inter-beam matrix, during or post acquisition
Threshold
Any single parameter or logical combination of multiple parameters
Data management
Workstation specifications (minimum required)
Clock speed of 3.2 GHz 16 GB RAM Hard drive and data storage
1 TB Solid State HD
Operating system Microsoft® Windows® 10 IoT 64-bit OS
Peripheral devices
At least 3 USB ports HP USB Keyboard US HP USB Optical Mouse
Networking
Ethernet LAN 10/100/1000
Signal Processing
18-bit dynamic range with IEEE 32 bit floating-point resolution
Monitor
LCD flat panel, 23 in.
LCD flat panel, 29 in. (recommended)
Data management options
BD FACSLinkTM software for LIS connectivity.
BD Assurity LincTM software for remote diagnostic capability
Installation requirements
Operating temperature
15°C (59°F) to 30°C (86°F)
Maximum of ±2.5°C/day fluctuation recommended
Humidity
15% to 85% relative humidity (noncondensing)
Dimensions (W x D x H)
Cytometer
63.2 x 57.9 x 57.9 cm
24.9 x 22.8 x 22.8 in.
With standard tanks
85.2 x 57.9 x 57.9 cm
33.5 x 22.8 x 22.8 in.
With standard tanks and loader 107.2 x 57.9 x 57.9 cm
42.2 x 22.8 x 22.8 in.
Weight
Cytometer: 56.0 kg (123.5 lb) Loader: 13.2 kg (29 lb)
Power specifications
Voltage: 100–240 ±10% VAC Frequency: 50–60 ±10% Hz Current: 2 A
Power: 200 W
Operational heat dissipation 488 BTU/hour at ambient temperature
Noise under normal operating conditions
55 dBA over 8 hours under normal operating conditions
Altitude
0.8 atm (approximately 2,000 meters)
System options
BD FACSTM Universal Loader Compatible with 30 (barcoded) or 40
(non-barcoded) tubes (12 x 75 mm). Equipped with an orbital shaker for in- place mixing and resuspension of cells. Optimized for all supported plate and tube formats. Includes internal barcode reader for positive sample identification.
Supported barcode formats Codabar Code 128
Code 3 of 9
Interleaved 2 of 5
Handheld barcode scanner
Handheld barcode scanner with stand supporting common 1-D and 2-D formats
Extended-use fluidics
Optional tanks and connectors to allow for use with 10-L waste tanks and
BD FACSFlowTM cubitainers
|